A. Features ;
1. Centrifugal concentrator with high
capacity cold trap cooler 2. Safety concentration without boiling over,
resist against acid and alkali 3. Simultaneous process of multiple
sample 4. Set parameter up to 60¡É for concentrating solvent. 5. Equipped
with vacuum gauge and safety device 6. Sealless vacuum by use of magnetic
driver 7. Use filter-attached leak path 8. System configuration 1)
Centrifugal Concentrator 1 set 2) Vacuum Pump 1 set 3) Freeze Trap 1
set
B. Specifications ;
1. Concentrator (VC-36N) 1 set 1)
Temperature range : Amb +5¢¦+60¡É 2) Capacity ( Simultaneous concentration ) :
1.5 ml microcentrifuge 36 ea 1.3 mm e.d. tube 16 ea 17.5 mm e.d. tube 8
ea 2) Temp. controller : Thermister on-off control 3) Heater : 90 W or
more 4) Spin speed : 120r/min~2000 r/min 5) Rotor spin method : Magnetic
drive method 6) Safety device : Over heat protect device Motorable
Key 7) Dimension : 310 x 340 x 377H mm 8) Aspirator path size : 14mm 9)
Weight : 24Kg 10) Power supply : AC100V, 2.5A
2 Vacuum Pump ( GLD-135) 1 set 1) Pumping speed :
200l/min or more 2) Pressure : 5x10-4 or more 3) Oil capacity : 600cc or
more 4) Weight : 23.5 Kg 5) dimension : 150x443x256
3. Freeze Trap ( VA-250F ) 1) Operation Temp. range : 0oC to
-50oC or more 2) Glass condenser capacity : 200 ml 3) Freezer : Closed
air freezing system, 150W 2) Cooling capacity : 100 Kcal/h or more 3)
Tank capacity : ¥õ135mm×210mm (3.0L) or more 4) Dimension : 325x435x540
mm
4. Glass Condenser (GC-035 )
C. System Configuration
1. Concentrator ( VC-36N ) 1 set 2. Freeze Trap ( VA-250F ) 1
set 3. Vacuum Pump ( GLD-135) 1 set 4. Glass Condenser ( GC-035 ) 1
set
D. Accessories
1. Rotor ( 36C ) 1 set 2. Vacuum
hose 2 meters
E. Remark
1. The installation and operation & test-run should be
carried out by the bidder at end-user's site. 2. Warranty should be guaranted
for one year.
|